Maintenance
This is a critical instrument for incident prevention and safety.
To maintain the performance of the detector and improve its reliability, perform regular maintenance.
NOTE
To use the pyrolyzer unit (PLU-70), also refer to the individual operating manual.
10-1
Maintenance intervals and items
•
Daily maintenance: perform maintenance before beginning to work.
•
Monthly maintenance: perform maintenance on the alarm circuit (alarm test) once a month.
•
Regular maintenance: perform maintenance once every six months.
Task
Description
Daily
Monthly
Biannually
Power supply
check
Verify that the power lamp lights up.
○
○
○
Concentration
display check
Verify that the gas reading is zero (or
20.9 vol% for oxygen). If the reading is
not zero (or 20.9 vol%), perform a zero
adjustment.
○
○
○
Flow rate
check
See the flow rate indicator to check for
abnormalities.
○
○
○
Filter check
Check the dust filter for dust or clogging.
○
○
○
Alarm test
Inspect the alarm circuit by using the
alarm test function.
○
○
Span
adjustment
Perform a span adjustment using
calibration gas.
○
Gas alarm
check
Check alarm operation using calibration
gas.
○
10-2
Calibration Overview
Calibration instructions depend on the GD-70D you are calibrating. The table below will tell you what
calibration instructions to use based on your GD-70D’s detection gas and range.
Detection Gas
Range
Calibration Instructions
AsH3 (Arsine)
0 - 0.20 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
B2H6 (Diborane)
0 - 0.30 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
BCl3 (Boron trichloride)
0 - 15.0 ppm
Calibrating with a Gas Bag (Calibrating with HCl) on page 99