- 87 -
GD-70D
Detection Gas
Range
Calibration Instructions
Br2 (Bromine)
0 - 1.00 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
BTBAS (Bis[tertiary-butyl-
amino]silane)
0 - 15.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
C2H2 (Acetylene)
0 - 2,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
C2H2Cl2 (Dichloroethene)
0 - 600 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
C2H4 (Ethylene)
0 - 2,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
C2H5OH (Ethanol)
0 - 2,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
C3H6 (Propylene)
0 - 5,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
C3H8 (Propane)
0 - 2,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
0 - 5,000 ppm
0 - 100% LEL
Calibrating with on page 90
C4F6 (Hexafluorabutadiene)
0 - 200 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
C4H10 (Isobutane)
0 - 100 %LEL
Calibrating with a Demand Flow Regulator on page 90
C6H14 (Hexane)
0 - 2,000 ppm
Calibrating with on page 90
CH2Cl2 (Dichloromethane)
0 - 2,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
CH3CN (Acetonitrile)
0 - 2,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
CH3COOH (Acetic acid)
0 - 30.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
CH3NH2 (Methylamine)
0 - 15.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
CH3OH (Methanol)
0 - 1,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
0 - 2,000 ppm
CH4 (Methane)
0 - 100 %LEL
Calibrating with a Demand Flow Regulator on page 90
0 - 2% vol
Calibrating with a Humidifier (MOS Sensors) on page 103
0 - 2000 ppm
0 - 5000 ppm
CH6Si (Methyl silane)
0 - 20.0 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
Cl2 (Chlorine)
0 - 3.00 ppm
Calibrating with a Demand Flow Regulator on page 90
ClF3 (Chlorine trifluoride)
0 - 0.60 ppm
Calibrating with a Surrogate Gas (Sensors with Factors) on
page 93
CO (Carbon monoxide)
0 - 75.0 ppm
Calibrating with a Demand Flow Regulator on page 90
0 - 150 ppm
0 - 300 ppm
COS (Carbonyl sulfide)
0 - 2,000 ppm
Calibrating with a Humidifier (MOS Sensors) on page 103
0 - 90.0 ppm
Calibrating a Pyrolyzer Unit on page 107