GD-70D
- 68 -
2-3.LAST CAL
Show the most recent
calibration date.
●
○
○
○
2- 3
LAST CAL
MAINTENANCE
→
SET
12:00
2009.01.01
MAINTENANCE
▲
↓
↑
▼
2-4.BIAS
Show the bias voltage "
BIAS
"
of electrochemical type (ESU)
or the element voltage
"
EVOLT
" of semiconductor
type (SGU) and hot-wire
semiconductor type (SHU).
●
○
○
○
2- 4
BIAS
MAINTENANCE
→
SET
0 mV
BIAS
MAINTENANCE
▲
↓
↑
▼
2-5.DEF FLOW
Set the flow rate to the default
value when the flow sensor is
replaced, etc.
Normally, this is not used
because the detector has
already been adjusted.
On the other hand, if the flow
meter is set at other than the
specified flow rate, gases
cannot be detected properly
due to an incorrect flow rate.
(Press the MODE key to
cancel this menu.)
●
○
○
○
2- 5
DEF FLOW
MAINTENANCE
⇔
SET
1000
DEF FLOW
MAINTENANCE
▲
↓
↑
▼
2-6.FLOW
Show the output and flow rate
of the current pump.
●
○
○
○
2- 6
FLOW
MAINTENANCE
⇔
SET
50 %
0.50 L/M
MAINTENANCE
▲
↓
↑
▼
2-7.TEMP
Show the temperature of the
detector.
●
○
○
○
2- 7
TEMP
MAINTENANCE
→
SET
25.0
°
C
TEMP
MAINTENANCE
▲
↓
↑
▼
2-8. WARMTIME
Show the suggested warm-up
completion date/time for
semiconductor type (SGU),
and hot-wire semiconductor
type (SHU) which requires a
long warm-up time.
●
○
○
○
2- 8
WARMTIM
E
MAINTENANCE
→
SET
12:00
2009.01.01
MAINTENANCE
▲
↓
↑
▼
2-9. SETTING1
Enter the environmental
setting 1 menu.
●
○
○
○
2- 9
SETTING1
MAINTENANCE
→
SET
Environmental
Setting 1
⇒
▲
↓
↑
▼