GD-70D
- 82 -
First Alarm Contact
Setting
Select either
nd
(De-
Energized)/
nE
(Energized),
and then press the SET key
to confirm the selection.
(Press the MODE key to
skip this menu.)
Factory setting is
nd
.
●
○
○
○
nd
AL1 RLY
MAINTENANCE
⇔
▲
/
▼
nE
AL1 RLY
MAINTENANCE
↓
↓
Second Alarm Contact
Setting
Select either
nd
(De-
Energized)/
nE
(Energized),
and then press the SET key
to confirm the selection.
(Press the MODE key to
skip this menu.)
Factory setting is
nd
.
●
○
○
○
nd
AL2 RLY
MAINTENANCE
⇔
▲
/
▼
nE
AL2 RLY
MAINTENANCE
↓
↓
Fault Alarm Contact
Setting
Select either
nd
(De-
Energized)/
nE
(Energized),
and then press the SET key
to confirm the selection.
(Press the MODE key to
skip this menu.)
Factory setting is
nd
.
●
○
○
○
nd
FLT RLY
MAINTENANCE
⇔
▲
/
▼
nE
FLT RLY
MAINTENANCE
↓
Return to
SET 6. RLY
PTRN
NOTE
When De-energized (
nd
) is selected, the relay is energized and activated during an alarm condition (de-
energized during normal operation).
•
When the alarm contacts are configured as normally open, the contacts are open during normal
operation and closed during an alarm condition.
•
When the alarm contacts are configured as normally closed, the contacts are closed during normal
operation and open during an alarm condition.
When Energized (
nE
) is selected, the relay is energized during normal operation (de-energized during an
alarm condition).
•
When the alarm contacts are configured as normally open, the contacts are closed during normal
operation and open during an alarm condition. They close if the power is turned OFF.
•
When the alarm contacts are configured as normally closed, the contacts are open during normal
operation and closed during an alarm condition.