41
Menu Operations and Setting Items
HPF
High-pass fi lter frequency setting
Enables a high-pass fi lter for the input signal. The HPF frequency indi-
cates the cutoff frequency.
Available settings are OFF and 5 Hz.
When the Inp setting is DC, only the OFF setting is available for the
high-pass fi lter.
OFF / 5 Hz
LPF
Low-pass fi lter frequency setting
Enables a low-pass fi lter for the input signal. The LPF frequency indicates
the cutoff frequency.
Available settings are OFF, 200 Hz, 1 kHz, and 2 kHz, but only selections
that are within the frequency range setting are allowed.
OFF / 200 Hz / 1 kHz / 2 kHz
Available parameter list
Frequency range
100 Hz 500 Hz 1 kHz 5 kHz 10 kHz 20 kHz
LPF
200 Hz
×
1 kHz
×
×
2 kHz
×
×
×
• • • Available
×
• • • Not available