73
Preparations
Sampling
The sampling action of the DA-21 is controlled by the frequency range and
sampling frequency setting. The sampling frequency can be set to 2.4 times
or 2.56 times the frequency range.
Setting the frequency range
The frequency range can be set in six steps (100 Hz, 500 Hz, 1 kHz, 5 kHz,
10 kHz, 20 kHz). Make the setting using the “Frequency Range” item in
<Rec.Parameters> menu.
The frequency range value represents the highest effective frequency that
will be included in the recorded waveform. Higher components will be cut
off. When making the setting, choose a value that is higher than the highest
frequency of components that need to be included in the recorded data.
Setting the sampling frequency
Waveform sampling is carried out at a frequency that is 2.4 times or 2.56
times the frequency range value. Make the setting using the “Sampling Fre-
quency” item in <Rec.Parameters> menu.
To perform FFT analysis after recording, using the ×2.56 setting is recom-
mended.
Содержание DA-21
Страница 2: ......
Страница 171: ...159 Specifications Unit mm Dimensional Drawings 175 45 140 Front view Right side view Rear view...
Страница 172: ...No 59374 16 07...