106
Recording
Recording parameters
Recording parameters
Recording parameters
Item
Setting
Item
Setting
Trigger mode
Repeat
Trigger type
External
Recording time
15 s
Pre Recording Time 5 s
Frequency range
20 kHz
Sampling frequency ×2.56
Input range
Suitable value
Procedure
1. Record calibration signal of sound level meter (at this time, trigger
mode should be set to “Free”).
2. Cancel calibration mode at sound level meter. Set trigger mode back
to “Repeat” and recording time to “15 s”.
3. Press the [REC] key to enable trigger standby.
4. Record voice memo as necessary.
5. Press the [STOP] key to terminate recording.
Note
It is also possible to select “SLM” for the “Sens” item
and set the sensitivity to the level range (full-scale
value) of the sound level meter, without recording
the calibration signal. However, for optimum mea-
surement accuracy, recording the calibration signal
of the sound level meter is preferable.