RIGOL
2-212
DS4000E Programming Guide
:TRIGger:RUNT
Command List:
:TRIGger:RUNT:ALEVel
Syntax
:TRIGger:RUNT:ALEVel <level>
:TRIGger:RUNT:ALEVel?
Description
Sets or queries the upper limit of trigger level of Runt trigger (namely the high level). The unit is the same
as the current amplitude unit of the source channel.
Parameter
Name
Type
Range
Default
<level>
Real
The range of <level> is from (-5.98 x VerticalScale -
Offset) to (6 x VerticalScale - Offset) (relative to the
vertical center of the screen).
Wherein, VerticalScale is the vertical scale of the source
channel, refer to the :CHANnel<n>:SCALe command;
Offset is the vertical position of the source channel, refer
to the :CHANnel<n>:OFFSet command.
1
Remarks
The upper limit of the trigger level of Runt trigger must be greater than the lower limit of the current trigger
level (refer to the :TRIGger:RUNT:BLEVel command).
Return Format
The query returns the upper limit of the trigger level in scientific notation.
:TRIGger:RUNT:BLEVel
Syntax
:TRIGger:RUNT:BLEVel <level>
:TRIGger:RUNT:BLEVel?
Description
Sets or queries the lower limit of trigger level of Runt trigger (namely the low level). The unit is the same as
the current amplitude unit of the source channel.