RIGOL
DS4000E Programming Guide
2-111
Parameter
Name
Type
Range
Default
<bool>
Bool
{{1|ON}|{0|OFF}}
0|OFF
Remarks
The display of the statistic information of the pass/fail test results can only be turned on when the pass/fail
test function is enabled (refer to the :MASK:ENABle command).
Return Format
The query returns 1 or 0.
:MASK:OPERate
Syntax
:MASK:OPERate <oper>
:MASK:OPERate?
Description
Runs or stops the pass/fail test operation.
Parameter
Name
Type
Range
Default
<oper>
Discrete
{RUN|STOP}
STOP
Remarks
You can only run the pass/fail test operation when the pass/fail test function is enabled (refer to
the :MASK:ENABle command).
Return Format
The query returns RUN or STOP.
:MASK:OUTPut
Syntax
:MASK:OUTPut <type>
:MASK:OUTPut?
Description
Enables or disables the alarm function; or queries the on/off status of the alarm function.
Parameter
Name
Type
Range
Default
<type>
Discrete
{FAIL|FSOund}
FAIL
Remarks
FAIL: When a failed waveform is detected, the oscilloscope displays the statistic results and outputs a
pulse; but the beeper does not sound.
FSOund: When a failed waveform is detected, the oscilloscope displays the statistic results and outputs
a pulse; while at the same time, the beeper sounds regardless of the on/off status (refer to
the :SYSTem:BEEPer command) of the beeper.
Return Format
The query returns FAIL or FSO.