RIGOL
2-154
DS4000E Programming Guide
Remarks
EXTRemum: The current value, average value, minimum value, and maximum value are displayed at
the bottom of the screen.
DIFFerence: The current value, average value, standard deviation, and count value are displayed at
the bottom of the screen.
Return Format
The query returns EXTR or DIFF.
:MEASure:STATistic:RESet
Syntax
:MEASure:STATistic:RESet
Description
Clears the history statistic data, makes statistics again, and displays the results.
:MEASure:SOURce
Syntax
:MEASure:SOURce <source>
:MEASure:SOURce?
Description
Sets or queries the source channel of the time, voltage, and area parameter measurement.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH}
CHANnel1
Remarks
When the math operation type is FFT (refer to the :CALCulate:MODE command), <source> cannot be
set to "MATH".
You need to specify two source channels for delay and phase parameter measurements. For the
source channel settings of delay parameter measurements, refer to the :MEASure:SETup:DSA
and :MEASure:SETup:DSB commands. For the source channel settings of phase parameter
measurements, refer to the :MEASure:SETup:PSA and :MEASure:SETup:PSB commands.
Return Format
The query returns CHAN1, CHAN2, CHAN3, CHAN4, or MATH.