RIGOL
DS4000E Programming Guide
2-163
:MEASure:VPP:SAVerage
:MEASure:VPP:SCURrent
:MEASure:VPP:SDEViation
:MEASure:VPP:SMAXimum
:MEASure:VPP:SMINimum
Syntax 1
:MEASure:VPP:SAVerage [<chan>]
:MEASure:VPP:SCURrent [<chan>]
:MEASure:VPP:SDEViation [<chan>]
:MEASure:VPP:SMAXimum [<chan>]
:MEASure:VPP:SMINimum [<chan>]
Description 1
Measures the peak-peak value of the specified source channel and enables the statistic function of the
measurement results.
Syntax 2
:MEASure:VPP:SAVerage? [<chan>]
:MEASure:VPP:SCURrent? [<chan>]
:MEASure:VPP:SDEViation? [<chan>]
:MEASure:VPP:SMAXimum? [<chan>]
:MEASure:VPP:SMINimum? [<chan>]
Description 2
Queries the average value, current value, standard deviation, maximum value, or minimum value of the
peak-peak value measurement results when the peak-peak value measurement function of the specified
source channel and the statistic function of the measurement results are enabled. The unit is related to the
current amplitude unit of the specified source channel.
Parameter
Name
Type
Range
Default
<chan>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH}
--
Remarks
Before making measurements, turn on the channels to be measured.
When the math operation type is FFT (refer to the :CALCulate:MODE command), <chan> cannot be
set to "MATH".
When [<chan>] is omitted, the command, by default, applies to the currently set measurement source
(refer to the :MEASure:SOURce command).
Return Format
The query returns the average value, current value, standard deviation, maximum value, or minimum value
in scientific notation. When the measurement results are invalid, the query returns 9.9e37.