Remarks
You do not need to set the data direction when the address width of IIC trigger is 8 bits (refer to
the :TRIGger:IIC:AWIDth command).
Return Format
The query returns READ, WRIT, or RWR.
:TRIGger:IIC:DLEVel
Syntax
:TRIGger:IIC:DLEVel <level>
:TRIGger:IIC:DLEVel?
Description
Sets or queries the trigger level of the data signal of IIC trigger. The unit is the same as the current
amplitude unit of the source channel.
Parameter
Name
Type
Range
Default
<level>
Real
The range of <level> is from (-6 x VerticalScale - Offset)
to (6 x VerticalScale - Offset) (relative to the vertical
center of the screen).
Wherein, VerticalScale is the vertical scale of the data line
source channel, refer to the :CHANnel<n>:SCALe
command; Offset is the vertical position of the data line
source channel, refer to the :CHANnel<n>:OFFSet
command.
0
Return Format
The query returns the trigger level in scientific notation.
:TRIGger:IIC:SCL
Syntax
:TRIGger:IIC:SCL <source>
:TRIGger:IIC:SCL?
Description
Sets or queries the source channel of the clock line of IIC trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Remarks
Channels that are not currently turned on can also be selected as the trigger source.
Return Format
The query returns CHAN1, CHAN2, CHAN3, or CHAN4.