![Rigol DS4000E Series Скачать руководство пользователя страница 144](http://html1.mh-extra.com/html/rigol/ds4000e-series/ds4000e-series_programming-manual_1460996144.webp)
RIGOL
2-132
DS4000E Programming Guide
:MEASure:NDUTy
Syntax 1
:MEASure:NDUTy [<chan>]
Description 1
Measures the negative duty cycle of the specified source channel.
Syntax 2
:MEASure:NDUTy? [<chan>]
Description 2
Measures and queries the negative duty cycle of the specified source channel.
Parameter
Name
Type
Range
Default
<chan>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH}
--
Remarks
Before making measurements, turn on the channels to be measured.
When the math operation type is FFT (refer to the :CALCulate:MODE command), <chan> cannot be
set to "MATH".
When [<chan>] is omitted, the command, by default, applies to the currently set measurement source
(refer to the :MEASure:SOURce command).
Return Format
The query returns the measurement results in scientific notation. When the measurement results are
invalid, the query returns 9.9e37.
:MEASure:NDUTy:SAVerage
:MEASure:NDUTy:SCURrent
:MEASure:NDUTy:SDEViation
:MEASure:NDUTy:SMAXimum
:MEASure:NDUTy:SMINimum
Syntax 1
:MEASure:NDUTy:SAVerage [<chan>]
:MEASure:NDUTy:SCURrent [<chan>]
:MEASure:NDUTy:SDEViation [<chan>]
:MEASure:NDUTy:SMAXimum [<chan>]
:MEASure:NDUTy:SMINimum [<chan>]
Description 1
Measures the negative duty cycle of the specified source channel and enables the statistic function of the
measurement results.
Syntax 2
:MEASure:NDUTy:SAVerage? [<chan>]
:MEASure:NDUTy:SCURrent? [<chan>]
:MEASure:NDUTy:SDEViation? [<chan>]
:MEASure:NDUTy:SMAXimum? [<chan>]
:MEASure:NDUTy:SMINimum? [<chan>]