Virtui
_____________
4. What communication method is being used?
RS-232___________
RS-485___________
Fiber–optic_________
R/F______________
5. What is the
iQUBE
configuration?
6. Do you have the manual for all the equipment in use?
7. After connection, is there Voltage at the
iQUBE
?
8. Is D1 On or Off on the connector board?
9.
If using a
920i
, continue to step 10.
If using
Revolution III
, access cell monitoring Via the interactive tab and proceed to step 18
If using
Virtui
, access cell monitoring and proceed to step 18
10. What is the distance from the
920i
to the scale?
11. What port is the
iQUBE
on?
If using port 3, ensure printer ports are not still set to the default port 3.
12. With the communications on, using
TE
S
TCOM
, are D5 and D6 or D7 and D8 flashing?
13. If using
iRev
, is the
CONFIG
already set up and ready to download?
14. Set up the diagnostic key.
15. Does the download complete to the
920i
without error?
16. Did the
920i
send the data to the
iQUBE
?
17. Does the Cal-Match work?
18. When the scale is empty, are all the current readings near the captured zero?
If yes, go to step 19.
If no, go to step 20.
19. With weight on the scale, do all of the current readings rise to the same degree?
If yes, go to step 20.
If no, go to step 21.