Rice Lake iQube 920i Скачать руководство пользователя страница 48

Appendix

45

9.0

Appendix

9.1

Troubleshooting

The following section provides information to assist
for troubleshooting or repairing the 

iQUBE

.

Materials needed to accomplish these steps are:

Revolution III

 loaded onto a laptop PC. If using

Virtui

iRev

 must be loaded on the PC

Digital voltmeter

9.1.1

General Steps

To determine that the 

iQUBE

 has power, check LED

D1 on the core module and LEDs for the load cells
assigned on each connector board (see Figure 2-2 on
page 4). If no load cell LEDs (D18–D25) are on,
download configuration to the 

iQUBE

.

Use a voltmeter to measure the voltage at connector
J10. Pin 1 should show 6 volts.

Verify communication LEDs D5 and D6 on the core
module are flashing (see Figure 2-2 on page 4). If
both LEDs are not flashing, check wiring and setup.

For 

Virtui

 applications, if the power and the LEDs are

all properly lit, use the 

Diagnostic

 softkey to go into

diagnostic mode. (This feature can be enabled in the
Feature menu in Config mode.) Use the diagnostic
displays to view the outputs of each load cell.

If no 

Virtui

 is available, connect the PC running

Revolution III

 to connector J7 with switch 6 in the off

position. After connecting, upload the 

iQUBE

c o n f i g u r a t i o n   t o   t h e   P C   t o   v e r i f y   t h e   p r o p e r
configuration is still loaded. If not, download the
correct configuration to the 

iQUBE

 and recalibrate.

Also verify that the 

iQUBE

 software is the latest

version.

If the configuration looks good, check the outputs of
each load cell. Cell outputs should change when a
weight on top of each cell or when the load cell cable
is disconnected.

Use the interactive menu of 

Revolution III

 to monitor

the cells, platforms, or systems. If any of the load cells
reads incorrectly, correct the problem then test the
total system again.

If these steps does not resolve the problem, contact
Rice Lake Weighing Systems technical support for
further assistance.

Symptom

Cause/Remedy

iQUBE

 seems dead

Verify power indicator LED on connector board is green.

Cycle power by pulling the 

iQUBE

 fuse and watch D1 on 

core module cycle through red, green, then off. If it does not, 
turn all switches off and only turn on 8. Cycle power again. 
This will clear the core module.

No LEDs lit for load cells

Download from 

Virtui

 required after setup is completed.

Some LEDs lit green or red

Green light means good config and good cell; red indicates a 
load cell fault or bad A/D conversion. Swap load cell cable to 
another position.

No scale or 

iQUBE

 in 

Virtui

 configuration

Configuration setup and download needed to setup 
configuration the 

iQUBE

Ensure setup switch (DIP switch 4 on the core module) is ON 
to enable download from 

Virtui

 to 

iQUBE

.

Print button causes drift

Change print format in PFORMT to send print to a port not 
connected to the 

iQUBE

.

Slaves not configured

Cycle power on primary unit only. Secondary units will receive 
new configuration data. 

Table 9-1. 

iQUBE

 Troubleshooting

Содержание iQube 920i

Страница 1: ...77224 Digital Diagnostic Junction Box Version 1 4 Installation Manual...

Страница 2: ...onnection to the 920i 11 2 11 2 Fiber Optic and Ethernet Communications 11 2 11 3 Remote Fiber Optic Module PN 77789 12 2 11 4 Analog Output 13 2 12 Installing Tap and Run Connectors for 920i Power 14...

Страница 3: ...c Tests 34 6 4 Diagnostic Setup 36 6 4 1 Configuring the 920i Diagnostics Softkey 36 6 4 2 920i Diagnostic Display 36 6 5 Error Messages 36 7 0 Alerts 38 7 1 Network Alert Notification 38 7 2 Alert Fo...

Страница 4: ...ital serial output Up to four iQUBE units can connect and communicate as a multidrop RS 485 network consisting of one primary and up to three secondary iQUBE units The maximum number of load cells sup...

Страница 5: ...ad cell outputs This is based on the reference established through cell associations The diagnostic conditions that can be identified for paired configurations are open bridge or channel zero referenc...

Страница 6: ...e 2 4 Load Cells When wiring load cells cells wired to iQUBE connectors J12 J19 8 channel units are assigned default names A1 A8 8 channel primary unit or B1 B8 8 channel secondary unit Be aware that...

Страница 7: ...SHLD 1 J16 LOAD CELL 5 SIG SIG EXC EXC SHLD 1 J17 LOAD CELL 6 SIG SIG EXC EXC SHLD 1 J18 LOAD CELL 7 SIG SIG EXC EXC SHLD 1 J19 LOAD CELL 8 SIG SIG EXC EXC SHLD 1 J21 1 IO 8 IO 7 IO 6 IO 5 IO 4 IO 3 I...

Страница 8: ...is only supported on Port A of the 2 channel expansion board Communications Cable Distance Limitations The maximum cable lengths that can be used for various communications types depend on a number of...

Страница 9: ...1 3 to set the secondary address as shown in Table 2 5 See Section 2 9 on page 10 for more information about the core module DIP switches Figure 2 3 920i Wiring DIP Switch Primary Unit Secondary Unit...

Страница 10: ...120 NOTE ALL SIX WIRES INCLUDING SHIELD AND GROUND WIRES MUST BE CONNECTED FOR COMMUNICATION BETWEEN iQUBE UNITS 120 J4 GND SHLD Z B Y A J7 GND SHLD Y B Z TXD A RXD 1 1 PRIMARY iQUBE UNIT J4 GND SHLD...

Страница 11: ...k rather than source switching current Each output is a normally open collector circuit capable of sinking 25 mA when active Digital outputs are wired to switch relays when the digital output is activ...

Страница 12: ...AC cards connect to J4 on the iQUBE core module See Figure 2 6 for J4 location NOTE Analog output setup information is supplied with the 920i VIRTUi or Revolution III technical documentation Analog ou...

Страница 13: ...see Figure 2 2 on page 4 2 Set core module DIP switch 8 ON 3 Power on the iQUBE Allow LED D1 on the core module to cycle from red to green or green to red several times then off 4 Power off the iQUBE...

Страница 14: ...more than two iQUBE units must use the iQUBE s optional internal power supply Figure 2 8 iQUBE Copper Wire Connection to 920i Showing RS 232 and DC Power Connections 2 11 2 Fiber Optic and Ethernet C...

Страница 15: ...quires that the iQUBE be fitted with its internal fiber optic card see Figure 2 9 on page 11 both voltage and current based DAC cards available for the remote fiber optic module Figure 2 10 Remote Fib...

Страница 16: ...12 Analog output from the iQUBE requires that a digital to analog DAC card be installed on the iQUBE core module Both voltage and current based DAC cards are available See Section 2 14 1 on page 15 fo...

Страница 17: ...connector until latched 3 Insert tap wire 6V supply to iQUBE into connector Use inspection port to check wire position 4 Use crimping tool to drive u contact down flush with top of connector Close hi...

Страница 18: ...gned with the edge of the core module iQUBE or the fiber optic interface board remote fiber optic module Carefully align the DAC card connector with the board connector J4 on the core module J5 on the...

Страница 19: ...for installation information Figure 2 18 920i Internal Fiber Optic Option Card 2 15 Fuse Replacement Fuse F1 on the iQUBE connector board see Figure 2 2 on page 4 provides protection for power supplie...

Страница 20: ...power ground 1 4 20NC x 3 4 hex head 1 8 15655 Cable grip 3 8 NPT 9 9 15656 Lock nut 3 8 NPT 10 10 15628 Cord grips 1 2 in NPT 10 11 15630 Lock nuts 1 2 in NPT 10 12 91785 iQUBE enclosure 1 18 77248...

Страница 21: ...18 iQUBE Installation Manual Figure 2 19 iQUBE Enclosure Assembly 1 3 5 6 7 9 8 10 11 4 12...

Страница 22: ...Installation 19 Figure 2 20 iQUBE Assembly 21 18 19...

Страница 23: ...he most complete and efficient configuration interface for the 920i 3 2 920i iQUBE Menu Configuration With both the iQUBE and the indicator powered off connect Port 3 or higher of the indicator serial...

Страница 24: ...QUBE configuration only then press enter 3 Cycle power on the iQUBE to initialize the changed configuration 4 Press the Save and Exit softkey 5 Return the 920i to setup mode 6 Continue configuration o...

Страница 25: ...at is not also routed to port 3 Use the PORT parameter to change the print format serial port assignment 3 4 1 SERIAL Menu iQUBE Submenu Figure 3 2 SERIAL Menu iQUBE Submenu SCALES SERIAL FEATURE PFOR...

Страница 26: ...f cells must specify CIRCULAR Full diagnostic capability requires PAIRED section configuration See Section 2 3 on page 3 for information about load cell build order PLATFM List of cells and platforms...

Страница 27: ...on pairs when half of the load cell outputs are below the specified threshold value CELLBAL ENABLE THRH TOLR LOAD Cell balance tolerance diagnostic subparameters See Table 3 2 on page 26 The cell bala...

Страница 28: ...EF number THRH number THRH number DELAY 0 125000 25 900 CELLEMU DIGIO TESTCOM OFF ENABLE ON If ENABLE ON number THRH number TOLR number LOAD 0 200000 25 5000 000 OFF ENABLE ON If ENABLE ON CELLBAL LDR...

Страница 29: ...value mV A m v reference value at which a cell is presumed to be unloaded This m v threshold value is added to the calculated zero m v of each cell THRH 25 value percent Specifies a band above the ca...

Страница 30: ...125000 value mV Peak to peak tolerance noise limit Specifies a millivolt limit for noise allowed between stored measurements CELLEMU ENABLE ON OFF Cell emulation Specifies whether cell emulation can b...

Страница 31: ...s attached to the 920i using port 3 ensure that the serial print format is not also routed to port 3 Use the PORT parameter to change the print format serial port assignment Figure 3 4 PFORMT Menu SCA...

Страница 32: ...p switch to place the iQUBE in setup mode 4 Start the Revolution III program 4 1 1 Revolution III Configuration To configure the iQUBE using Revolution do the following 1 With the PC and iQUBE connect...

Страница 33: ...disk space Recommended system 233 MHz x86 compatible or greater with 64MB RAM 40 MB disk space iRev runs on most Windows operating systems including Windows 95 original release Windows 95 OSR2 Windows...

Страница 34: ...perform a Cal Match calibration of the iQUBE using the 920i front panel do the following 1 Place the indicator in setup mode display reads Scale Configuration and remove all weight from the scale pla...

Страница 35: ...value The displayed weight for the scale system is shown at the bottom of the weight column Tweaking can be performed following a Cal Match calibration or when replacing a load cell 5 1 4 Final Calibr...

Страница 36: ...You can view the active status of the cells platform or system by returning to the Interactive menu and selecting Status Monitoring for cells platforms or systems Click on the Auto Refresh box to auto...

Страница 37: ...ion 920i Default Revolution III iQUBE Default Zero Reference Which cell is out of tolerance at zero threshold the upper limit of the zero reading of the cells 0 125mv 0 200mv Per Cent threshold the pe...

Страница 38: ...to obtain error Multiple sections can error at the same time Threshold 125 m v 25 Threshold Calibrated Dead Load C1 C3 C4 C2 Threshold 125 m v 25 Threshold Calibrated Dead Load C1 C2 C3 C4 Zero Refer...

Страница 39: ...ighlighted cell Use the indicator navigation keys to select cells NOTE Displayed deadload and current values are raw millivolt values All diagnostic measurements use corrected mV values corrected_valu...

Страница 40: ...il Cell Over Under Range Weighing Error Error 2 Cell Long Term Drift Error Cell Balance Error Cell Noise Error Cell At Load Drift Error Zero Reference Cell Error Master Primary Unit Error Error 3 Slav...

Страница 41: ...Rev or by using the ALERT SERV serial command The SMTP server address cannot be set using the 920i menus To disable alert notification set the SMTP server address to 0 0 0 0 The alert contact e mail a...

Страница 42: ...ion 8 0 on page 42 or the iRev configuration utility See the 920i Installation Manual PN 67887 for more information about print formatting Command Content COMP Company name up to 30 characters COAR1 C...

Страница 43: ...mail_address Enter the email address of the contact company or dealer If the iQUBE alert support is used to send automated alert messages see Section 7 0 on page 38 enter the email address to which th...

Страница 44: ...to the 920i using port 3 ensure that the serial print format is not also routed to port 3 Use the PORT parameter to change the print format serial port assignment Figure 7 3 PFORMT Menu SCALES SERIAL...

Страница 45: ...y 1 500000 SJ LC SENS c n Load cell sensitivity 0 0001 9 9999 mV V SJ LC TRIM c n Load cell trim factor 0 0001 9 9999 SJ LC ZERO c n Load cell zero value 0 5 to 45 0 For load cell SJ LC commands endin...

Страница 46: ...diagnostic enable ON OFF SJ DIAG CB THRH n Cell balance threshold 0 0 9 999999 SJ DIAG CB TOLR n Cell balance percent tolerance 0 100 SJ DIAG CB LOAD n Cell balance load 0 0 9999999 0 SJ DIAG LDRIFT...

Страница 47: ...put full scale 0 000001 9999999 0 For commands ending with n n is the number of the 920i serial port to which the iQUBE is attached Table 8 7 ALGOUT Serial Commands Valid Only If Analog Output Option...

Страница 48: ...that the iQUBE software is the latest version If the configuration looks good check the outputs of each load cell Cell outputs should change when a weight on top of each cell or when the load cell ca...

Страница 49: ...M to verify communications This step retrieves the serial numbers of the attached iQUBE units for verification Check communication wiring and switches 5 and 6 No communication with secondary unit D7 a...

Страница 50: ...step 18 10 What is the distance from the 920i to the scale 11 What port is the iQUBE on If using port 3 ensure printer ports are not still set to the default port 3 12 With the communications on usin...

Страница 51: ...s broken or disconnected 24 Compared to the other cells is the single cell reading very low If yes the EXC wire is broken 25 Is the cell reading high low or not moving This could be a binding cell or...

Страница 52: ...the PC and that the cable from the programmer is plugged into the PC cable 1 Insert core board in the iQUBE grey box test fixture or production box 2 Remove the two jumpers on J3 3 Connect the blue c...

Страница 53: ...idity 0 95 relative humidity Enclosure Enclosure Dimensions 11 5 in x 8 25 in x 5 25 in 29 cm x 21 cm x 13 cm Weight 8 0 lb 4 Kg Rating Material NEMA 4X Certifications and Approvals NTEP CoC Number 03...

Страница 54: ...ccident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equipment has not been modified altered or...

Отзывы: