![Rice Lake iQube 920i Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/rice-lake/iqube-920i/iqube-920i_installation-manual_1453911012.webp)
Installation
9
2.8
Analog Output
The analog output option consists of either a 0-30mV (PN 77146) or 0-20mA (PN 77797) digital-to-analog
(DAC) card that plugs into the
iQUBE
core module. The 0–30 mV output simulates the output of a load cell or
standard junction box, allowing connection to standard indicators. A 0–20 mA (passive only) DAC is also
available for PLC applications. the 0-20mA and the 0-30mV DAC cards connect to J4 on the
iQUBE
core
module. See Figure 2-6 for J4 location.
NOTE
: Analog output setup information is supplied with the
920i
,
VIRTUi
, or
Revolution III
technical
documentation.
Analog output from the
iQUBE
is sent to connector J9 on the connector board (see Figure 2-2 on page 4). See
Section 2.14.1 on page 15 for more information about installing the analog output option.
Figure 2-6. 0-20mA Analog Output Wiring
Figure 2-7. 0 - 30mV and 0 - 20mA Converters
J9
ANALOG 1
+
S
IG
1
–
S
IG
S
HLD
ON
CORE MODULE
PLC
+IN
-IN
+24
+24V Ret
Analog Input
N104
Power Supply
150
DAC
Connection
J4