Standard 30 MF
110-700
Operator and Service Manual
Edition: 8
2-9
4361.DOC010302 AUG18.DOCX
The factory preset CAN-Bus address for the Operator Unit 130-627 is
“09”.
The factory preset CAN-Bus address for the Standard 30 MF is “17”.
After a change of CAN-Bus address a reset has to be performed.
2.3.7
Connection of the Operator Unit Standard 30 MF 130-627
During the setting to work procedure, the Operator Unit 130-627 has
to be connected to the cable clamps to make all the necessary
presettings. If the Standard 30 MF is used in a system environment
the Operator Unit 130-627 is disconnected again after completion of
the setting to work procedure and the Operator Unit 130-626 is used
for further operation.
In a standalone environment the Operator Unit 130-627 is also used
for operation.
If the Operator Unit 130-627 should be used for operation in a
standalone system it has to be mounted in a console. Information
about mounting and settings see dimensional drawing 130-
627.HP005 in the annex.
1. Adjust the configuration of the Operator Unit 130-627 to the given circumstances.
2. Connect the Operator Unit 130-627 to the cable terminals.