Preliminary
F.6 Oil Correction Table
In the Setup it is possible to make a measurement correction of the PF / tan δ on the basis of the
temperature.
To access this feature, you have to enter the service code “
1200#
” first.
In the Setup you have access to this menu.
Here you can activate it and choose the
correction table.
With Info you see the table on the screen.
Delete will delete the selected table.
Use Import / Export to exchange tables with the connected USB-key.
The file looks like IEEE C57.12.90.oct
. It is on the root of the USB-key. You can edit the file e.g.
with a text editor, Microsoft Excel, Libre Office Calc etc.
This is an Example of an Oil Correction
Table.
It is possible to add customer tables
tailored to the specific oil or test object.
1
The standar
td
table was from the IEEE C57.1290, you see the correction factor for the tan delta at the specific temperature
(between it' s an interpolation). This table is depending from the test object and it's not really an accurate correction. Now it's not any
more in this standard, but some customers still want to have it, that's why we use a service code for it.
90315-1.2 Instruction Manual CAPO
Page 107 of 117
Содержание CAPO 2.5
Страница 37: ...Preliminary 4 Thermal Printer 90315 1 2 Instruction Manual CAPO Page 37 of 117 ...
Страница 96: ...Appendix Page 96 of 117 CAPO 2 5 CAPO 12 90315 1 2 Instruction Manual CAPO CAPACITANCE AND POWER FACTOR METER ...
Страница 113: ...Preliminary I Warranty Conditions 90315 1 2 Instruction Manual CAPO Page 113 of 117 ...
Страница 117: ...Preliminary K Index Page 117 of 117 90315 1 2 Instruction Manual CAPO ...