Preliminary
1: Main Switch
Press to power the CAPO on / off.
2: Main Input
Connect the power cable to the Main Input.
Press the Main Switch to switch the CAPO on / off.
A slow acting Fuse (T 10A) protects the device. (Use only the correct fuse type to replace it).
3: Safety GND
Connect a proper earth.
The FIRST cable you have to connect and lead to earth ground is always the Safety GND. This is
also the LAST cable to disconnect! Any interruption of the grounding connection can create an
electric shock hazard.
90315-1.2 Instruction Manual CAPO
Page 36 of 117