Preliminary
H Trouble Shooting
When the CAPO is powered on, an internal calibration and check sequence is performed.
Upon completion of the check sequence, the test set will proceed to the “Main Screen”.
If there any problem detected with the test set, an err or message will appear.
Test Voltage cannot be turned on
Press the safety-switch before starting the measurement.
If the contact is open, the safety-switch is not
pressed.
The safety GND or the HV GND is not
connected
The earth-symbol is crossed if an earth is
missing.
Is the Emergency stop switch pushed in? Turn it
counter clockwise to release.
If the emergency button is pressed, there is a
mushroom button visible.
90315-1.2 Instruction Manual CAPO
Page 111 of 117