Preliminary
D.3.2 Asynchronous Answer
Events/Redirected Messages:
Response Code
Remarks
"@*10 Exc,"
Redirected Exception, e.g Over current
"@*11 Err,"
Redirected Error
"@*12 Wrn,"
Redirected Warning
"@*13 Msg"
Redirected Message, Status
"@*15 Settings,"
Redirected Settings Message
"@*19 Set to Local"
"Local" Button pressed on the CAPO
"@*20 Start"
Measurement started
"@*21 End"
Measurement stopped
"@*R0,"
Results formatted value with unit
"@*R1,"
Results formatted as float
D.3.3 Serial Interface
Interface Parameters
Parameter
Port
Baudrate
Databit
Stopbits
Parity
TxD (CAPO-> Host)
RxD (Host-> CAPO)
Gnd
Signal
Value
RS232
38400Baud
8 Bit
1 Bit
No
Pin 2
Pin 3
Pin 7
+/- 12V
D.4 USB Interface
To communicate with the CAPO over USB it is necessary to switch to Raytech Driver Mode.
Enter Service Code “2001” to set USB Driver Mode. This is normally already done. For more
information have a look at chapter F.4 Raytech USB Driver on page 106.
90315-1.2 Instruction Manual CAPO