Preliminary
8.2.6.1 Single measurement
To start a single measurement, press and
hold the Safety Switch and afterwards
press the Single button
The CAPO will do the measurement and
after you get a stable result, it will
automatically stop the measurement.
You can Save the result or directly Print
it on paper
8.2.6.2 Multi measurement
Instead of Single the button can be labeled
with Multi . This are multiple Single
measurements.
Mulit is visible if frequency- or voltage-
"Sweep" is selected or the measurement
Mode "Xfrmr" or "Multi".
There is more information in the following chapter:
8.2.4 Measurement frequency on page 49
8.2.5 Measurement voltage on page 51
9.5 Xfrmr on page 81
9.6 Multi on page 84
90315-1.2 Instruction Manual CAPO
Page 54 of 117