Preliminary
6.2.3.1 Simplifications for small Dissipation / Power Factor
If the dissipation / power factor is small, the error can be neglected and you can use the following
simplifications:
For tan δ < 3%, the difference between PF and tan δ is smaller than 0.1%
where:
Cs
serial Capacitance [F]
Cp
parallel Capacitance [F]
Z
Impedance [Ω]
Xs
serial reactance [Ω]
Xp
parallel reactance [Ω]
Rs
serial Resistance [Ω]
Rp
parallel Resistance [Ω]
90315-1.2 Instruction Manual CAPO
Page 31 of 117
Z
≈
Xs
≈
Xp
Cs
≈
Cp
Rs
≈
Xs
∗
tan
δ
Rp
≈
Xp
tan
δ
Error
≈
1
+
tan
2
δ
Содержание CAPO 2.5
Страница 37: ...Preliminary 4 Thermal Printer 90315 1 2 Instruction Manual CAPO Page 37 of 117 ...
Страница 96: ...Appendix Page 96 of 117 CAPO 2 5 CAPO 12 90315 1 2 Instruction Manual CAPO CAPACITANCE AND POWER FACTOR METER ...
Страница 113: ...Preliminary I Warranty Conditions 90315 1 2 Instruction Manual CAPO Page 113 of 117 ...
Страница 117: ...Preliminary K Index Page 117 of 117 90315 1 2 Instruction Manual CAPO ...