Figure 3-19
Vapor: Select a substance to calibrate on (pos)
13. Select a substance from the positive ion list of substances.
•
To calibrate using the internal calibrant, select
Vapor-Pos-CAL
.
•
To calibrate using the Check Sample vial, select
Vapor-Pos-Ver
.
•
To calibrate using one of the substances in the list, select the appropriate substance.
14. Press OK.
A message displays to select the correct peak.
Figure 3-20
Vapor: Vapor-Pos-CAL peak
15. Verify the calibration peak displays in the white window.
16. Touch slightly above the top of the peak until the dot, drift time, and peak height display.
CHAPTER 3 │User's Guide
39