Calibration Warning
The
Calibration Warning indicates that routine calibration needs to be performed. This warning displays every 8
hours, by default.
Figure 6-5
Calibration Warning
Resolution
1. Touch the warning on the status bar to display the warning details.
Figure 6-6
Calibration Warning details
2. Press Close.
3. Select Menu > Manual Calibrate, or press Cal to calibrate the device.
see "Calibration and Verification"
Heater Warning
The
Heater Warning indicates there could be a problem with the internal heater due to a short circuit in the
temperature sensor wiring or open wiring.
106
Rapiscan Systems Proprietary and Confidential
CHAPTER 6 │ User's Guide