The device automatically detects the vapor plate and adjusts the temperature for vapor sampling.
8. Log on to the ITMS™ software.
V (for Vapor) displays in the status screen. Allow approximately 20 minutes for the device to warm up to
the correct temperature.
Taking a Vapor Sample
1. Place the vapor plate as close to the object being sampled without touching, preferably around seams or
areas where vapors may seep out.
Figure 5-12
Vapor sampling
2. Press Trig, the green button on the navigation pad, or the green button on the device handle.
3. Collect a vapor sample for 8-10 seconds.
4. Press Stop to stop the collection process.