Specifications: Fault and Warning Messages
American Science and Engineering, Inc.
MINI
Z
Operator Manual 342-9000-1 Rev. B ECCN EAR99
A-5
Image export failed
[Warning]
The system was unable to convert the current image to JPG file format or was
unable to transfer the resulting JPG file to an external storage device. Verify that
the target storage device is connected and has sufficient free space to accept
the s.
Interlock fault
[Fault]
The interlock sensing mechanism has a fault, and the scanner X-ray generation
stages are shut down as a precaution. Turn the key off and then on again to
attempt to reset the scanner. If the fault condition persists, turn the key off and
contact AS&E.
Light leak
[Fault]
Ambient light is entering the detection elements, blanking the elements, and the
scanner X-ray generation stages are shut down as a precaution. Turn the key
off and then on again to attempt to reset the scanner. If the fault condition
persists, turn the key off and contact AS&E.
Motor control fault
[Fault]
The scanner motor controller has a fault, and the scanner X-ray generation
stages are shut down as a precaution. Turn the key off and then on again to
attempt to reset the scanner. If the fault condition persists, turn the key off and
contact AS&E.
Motor speed fault
[Fault]
The re are two speed tolerance thresholds for scanner motor speed. When the
scanner motor reaches the second limit, 20% either too fast or too slow, and a
scan is attempted, this fault message is displayed, and the scanner X-ray
generation stages are shut down as a precaution. Turn the key off and then on
again to attempt to reset the scanner. If the fault condition persists, turn the key
off and contact AS&E.
Motor speed too fast or too
slow
[Warning]
There are two speed tolerance thresholds for scanner motor speed. When the
scanner reaches the first limit, 10% either too fast or too slow, this warning
message is displayed (image quality may be degraded). This warning clears
when the scanner enters Standby mode.The user should suspend scanning
operation until the motor speed returns to the normal range.
PMT high voltage fault
[Fault]
A component in the X-Ray detection system has an error, and the scanner X-ray
generation stages are shut down as a precaution. Turn the key off and then on
again to attempt to reset the scanner. If the fault condition persists, turn the key
off and contact AS&E.
Power LED fault
[Fault]
Indicates that the Power LED (located on the scanner housing) is being
supplied by current that is either too high or too low, and the scanner X-ray
generation stages are shut down as a precaution. Turn the key off and then on
again to attempt to reset the scanner. If the fault condition persists, turn the key
off and contact AS&E.
Scanner battery low
[Warning]
There are two scanner battery thresholds for level of charge. When the scanner
battery charge has depleted to the first threshold, less than 15% of full charge,
this message is displayed. Turn the key off and replace the scanner battery with
a fully charged battery, then restart the scanner. Place the depleted battery in
the scanner battery recharger and check the battery status. Recharge the
battery if it is capable.
Table A-4: Fault and Warning Messages (continued)