Maintenance: Recalibrating the Scanner
American Science and Engineering, Inc.
MINI
Z
Operator Manual 342-9000-1 Rev. B ECCN EAR99
11-11
7. In the
Display
tab, tap the
Setup
button. The screen updates. The user is prompted to tap the
touchscreen to validate that the touchscreen is the screen to be recalibrated.
8. Tap the screen. The
Display
tab opens again. The
Display
option is
1. Internal LCD
. This is the
only choice.
9. Tap the
Calibrate
button. The utility presents a series of screens with a crosshairs target in each. In
each screen use the stylus to tap the center of the crosshairs target as accurately as possible.
10. Complete the procedure without changing orientation of the screen.
11. If the overall result is unsatisfactory, in the
Display
tab, tap the
Reset
button and try again.
11.8
Recalibrating the Scanner
Calibration of the MINI Z system is performed at the AS&E factory before shipment to the customer,
and typically the system does not require recalibration. In the event a significant drop or other physical
Содержание AS&E MINI Z