Configuration
R&S
®
ESW-K58
38
User Manual 1179.0880.02 ─ 03
While the measurement is running, the "Single Sweep" softkey and the [RUN SINGLE]
key are highlighted. The running measurement can be aborted by selecting the high-
lighted softkey or key again.
Note:
Sequencer. Furthermore, the [RUN SINGLE] key controls the Sequencer, not
individual sweeps. [RUN SINGLE] starts the Sequencer in single mode.
If the Sequencer is off, only the evaluation for the currently displayed channel is upda-
ted.
Remote command:
Measurement mode:
Run measurement:
5.10
Transducer
Many EMC test setups contain a transducer (for example antennas, cables, probes or
current probes). The transducer converts the interference variables like field strength,
current or RFI voltage into a voltage across 50
Ω
. Because most transducers have a
characteristic frequency response, it is necessary to correct the measurement results
by the frequency characteristics of the transducer. These characteristics are defined in
a transducer factor or transducer sets.
For more information on how to use transducers and how to work them, refer to the
R&S
ESW user manual.
Transducer