Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
71
User Manual 1179.0880.02 ─ 03
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Default unit: OHM
Example:
INP:IMP 75
Manual operation:
See
INPut<ip>:TYPE
<Input>
The command selects the input path.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Input>
INPUT1
Selects RF input 1.
INPUT2
Selects RF input 2.
*RST:
INPUT1
Example:
//Select input path
INP:TYPE INPUT1
Manual operation:
See
INPut<ip>:ATTenuation:LIMiter[:STATe]
<State>
This command turns the pulse limiter on and off.
The pulse limiter is an additional protection mechanism for the second RF input that
attenuates high level pulses.
Suffix:
<ip>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
ON
Example:
//Turn on pulse limiter
INP:ATT:LIM ON
Manual operation:
See
8.5.1.2
External mixer
The remote commands to configure external mixers are the same as in the Spectrum
application.
Configuration