Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
79
User Manual 1179.0880.02 ─ 03
Example:
FREQ:OFFS 1GHZ
Manual operation:
See
[SENSe:]FREQuency:TUNed
<TunedFrequency>
Sets/queries the tuned frequency in Hz.
Parameters:
<TunedFrequency>
Default unit: Hz
Example:
SENS:FREQ:TUNed 1 GHz
Manual operation:
See
8.5.5
Trigger configuration
................................................................................... 79
TRIGger<tp>[:SEQuence]:LEVel[:EXTernal]
........................................................................79
TRIGger<tp>[:SEQuence]:HOLDoff[:TIME]
......................................................................... 80
.............................................................................................80
......................................................................................81
TRIGger[:SEQuence]:IFPower:HOLDoff
............................................................................. 81
OUTPut<up>:TRIGger<tp>:DIRection
................................................................................ 81
TRIGger<tp>[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
Note on external triggers:
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure that this situation is avoided in your remote control programs.
Suffix:
<tp>
.
irrelevant
Parameters:
<Source>
See table below.
*RST:
IMMediate
Example:
//Select external trigger input as source of the trigger signal
TRIG:SOUR EXT
Manual operation:
See
See
See
TRIGger<tp>[:SEQuence]:LEVel[:EXTernal]
<Level>
This command defines the level the external signal must exceed to cause a trigger
event.
Configuration