Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
76
User Manual 1179.0880.02 ─ 03
Example:
//Turn on auto ranging
INP:ATT:AUTO ON
Manual operation:
See
INPut<ip>:ATTenuation:PROTection[:STATe]
<State>
This command turns the availability of attenuation levels of 10
dB or less on and off.
Suffix:
<ip>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
1
Example:
//Turn on input protection
INP:ATT:PROT ON
Manual operation:
See
INPut<ip>:ATTenuation[:VALue]
<Attenuation>
This command defines the attenuation at the RF input.
To protect the input mixer, attenuation levels of 10
dB or less are possible only if you
have turned off the input protection with
INPut<ip>:ATTenuation:PROTection[:
Suffix:
<ip>
.
irrelevant
Parameters:
<Attenuation>
Range:
0 dB to 79 dB
*RST:
10 dB
Default unit: dB
Example:
//Define attenuation
INP:ATT 40dB
Manual operation:
See
INPut<ip>:GAIN:AUTO
<State>
This command includes and excludes the preamplifier from the auto ranging feature.
Suffix:
<ip>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Configuration