Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
83
User Manual 1179.0880.02 ─ 03
[SENSe:]MAPD:NCHannels
<Value>
Sets or queries the number of channels that are analyzed.
Parameters:
<Value>
Number of channels
Example:
SENS:MAPD:NCH 21
Manual operation:
See
[SENSe:]FREQuency:SPAN
<Span>
This command defines the frequency span.
Parameters:
<Span>
Default unit: Hz
Example:
FREQ:SPAN 20MHZ
Defines a span of 20 MHz.
Manual operation:
See
[SENSe:]BANDwidth[:RESolution]
<Bandwidth>
Defines the resolution bandwidth and decouples the resolution bandwidth from the
span.
Parameters:
<Bandwidth>
<numeric value>
Range:
1 Hz to 1 MHz
Default unit: Hz
Example:
//Select resolution bandwidth
BAND:AUTO OFF
BAND 100KHZ
Manual operation:
See
[SENSe:]SWEep:TIME
<Time>
This command defines the measurement time. It automatically decouples the time from
any other settings.
Parameters:
<Time>
refer to data sheet
*RST:
depends on current settings (determined automati-
cally)
Default unit: S
Manual operation:
See