Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
91
User Manual 1179.0880.02 ─ 03
FORMat:DEXPort:DSEParator
<Separator>
This command selects the decimal separator for data exported in ASCII format.
Parameters:
<Separator>
POINt | COMMa
COMMa
Uses a comma as decimal separator, e.g.
4,05
.
POINt
Uses a point as decimal separator, e.g.
4.05
.
*RST:
*RST has no effect on the decimal separator.
Default is POINt.
Example:
FORM:DEXP:DSEP POIN
Sets the decimal point as separator.
Manual operation:
See
MMEMory:STORe<n>:TRACe
<Trace>, <FileName>
This command exports trace data from the specified window to an ASCII file.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
ESW base unit user manual.
Suffix:
<n>
.
Parameters:
<Trace>
Number of the trace to be stored
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR1:TRAC 1,'C:\TEST.ASC'
Stores trace 1 from window 1 in the file
TEST.ASC
.
8.6.2
APD display configuration
CALCulate<n>:MARKer<m>:Y:PERCent
............................................................................ 92
CALCulate<n>:SGRam:THReedim[:STATe]
........................................................................ 92
CALCulate<n>:SPECtrogram:THReedim[:STATe]
................................................................92
CALCulate<n>:SGRam:THReedim:LIMits
...........................................................................93
CALCulate<n>:SPECtrogram:THReedim:LIMits
.................................................................. 93
Analysis