Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
80
User Manual 1179.0880.02 ─ 03
Note that the variable [Input/Output] connectors must be set for use as input using the
OUTPut<up>:TRIGger<tp>:DIRection
command.
Suffix:
<tp>
.
irrelevant
Parameters:
<Level>
Default unit: V
Example:
//Define a trigger level of 2 V for an external trigger source
TRIG:SOUR EXT
TRIG:LEV 2V
Manual operation:
See
TRIGger<tp>[:SEQuence]:HOLDoff[:TIME]
<Offset>
Defines the time offset between the trigger event and the start of the measurement
(data capturing).
A negative offset is possible for time domain measurements.
For the trigger sources "External" or "IF Power", a common input signal is used for both
trigger and gate. Therefore, changes to the gate delay affect the trigger offset as well.
Suffix:
<tp>
.
irrelevant
Parameters:
<Offset>
For measurements in the frequency domain, the range is 0
s to
30
s.
For measurements in the time domain, the range is the negative
sweep time to 30
s.
*RST:
0 s
Default unit: s
Example:
//Define a trigger offset
TRIG:HOLD 500us
Manual operation:
See
TRIGger[:SEQuence]:DTIMe
<DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
Parameters:
<DropoutTime>
Dropout time of the trigger.
Range:
0 s to 10.0 s
*RST:
0 s
Default unit: S
Manual operation:
See
Configuration