Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
75
User Manual 1179.0880.02 ─ 03
<w>
subwindow
Not supported by all applications
<t>
irrelevant
Parameters:
<ReferenceLevel>
The unit is variable.
Range:
see datasheet
*RST:
0 dBm
Default unit: DBM
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
<Offset>
This command defines a reference level offset (for all traces in all windows).
Suffix:
<n>
.
irrelevant
<w>
subwindow
Not supported by all applications
<t>
irrelevant
Parameters:
<Offset>
Range:
-200 dB to 200 dB
*RST:
0dB
Default unit: DB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation:
See
"Shifting the Display (Offset)"
INPut<ip>:ATTenuation:AUTO
<State>
This command turns automatic determination of the attenuation level on and off.
When you turn it on, the R&S
ESW selects an attenuation that results in a good signal-
to-noise ratio without overloading the RF input.
Suffix:
<ip>
.
irrelevant
Parameters:
<State>
ON | OFF
ON
Selects automatic attenuation mode.
OFF
Selects manual attenuation mode.
*RST:
ON
Configuration