32
32-bit floating-point numbers
For I/Q data, 8 bytes per sample are returned for this format set-
ting.
64
64-bit floating-point numbers
Compared to
REAL,32
format, twice as many numbers are
returned.
Example:
FORM REAL,32
TRACe<n>[:DATA]?
<ResultType>
This command queries current trace data and measurement results.
The data format depends on
Suffix:
<n>
.
Query parameters:
<ResultType>
Selects the type of result to be returned.
TRACE1
Returns the trace data for trace 1 for Channel1 to ChannelM,
TracePoint1 to TracePointN.
The data are returned in the order Channel1/TracePoint1, ... ,
Channel1/TracePointN, Channel2/TracePoint1, ... , Channel2/
TracePointN, ... , ChannelM/TracePoint1, ... ChannelM/Trace-
PointN.
Channels that are not active return NAN values.
Example:
TRAC? TRACE3
Queries the data of trace 3.
Manual operation:
See
CALCulate<n>:STATistics:RESult<res>?
<ResultType>
This command queries the results of a measurement for a specific trace.
Suffix:
<n>
.
<res>
Query parameters:
<ResultType>
MEAN
Average (=RMS) power in dBm measured during the measure-
ment time.
PEAK
Peak power in dBm measured during the measurement time.