Configuration
R&S
®
ESW-K58
35
User Manual 1179.0880.02 ─ 03
Load Prede-
fined Settings
The button "Load Predefined Settings" opens the file load dialog with
the default directory of standard files:
C:\R_S\INSTR\mapd_std
.
You can select one of the standards that are stored here. The firm-
ware, in default case, provides the standard file
CISPR_11_Ed.6.2.xml
. The standard can only be loaded if the
Center Frequency is 1 GHz >= f <= 2,4 GHz, 2,5 GHz >= f <= 5,72
GHz and 5,88 GHz >= f <= 18 GHz.
Save Prede-
fined Settings
The button "Save Predefined Settings" saves the current settings
(Measurement Settings except Center Fre Evaluation Set-
tings) to a user-selected standard file. The default directory for the
save dialog is
C:\R_S\INSTR\mapd_std
.
Remote command:
Measurement Settings
The measurement settings area provides functionality to define the parameters of the
measurement.
If parameters in the measurement settings are changed in a way that they no longer
correspond to the evaluation table, the corresponding column in the table is grayed
out.
No of Chan-
nels
Defines the number of measurement channels.
Span
Defines the measurement span.
Analysis Band-
width
Defines the analysis bandwidth for the measurement.
Acquisition
Time
Defines the acquisition time for the measurement.
Generate Eval-
uation Table
from Measure-
ment Settings
Takes the user set center frequency, span and analysis bandwidth
and populates the evaluation table out of these values.
Adjust Center
Frequency to
not exceed
CISPR Band
Edges
Provides functionality to resolve CISPR band selection issues.
If the center frequency exceeds the CISPR band egdes, a yellow
warning message is displayed in the status bar.
Remote command:
[SENSe:]BANDwidth[:RESolution]
[SENSe:]MAPD:CHANnel<n>:POPulate
Measurement settings