Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
88
User Manual 1179.0880.02 ─ 03
Parameters:
<Value>
numeric value
Default unit: Hz
Example:
SENS:MAPD:CHAN1:RFR 1 MHZ
Manual operation:
See
[SENSe:]MAPD:CHANnel<n>:SAVE
<FileName>
Suffix:
<n>
.
1..n
Parameters:
<FileName>
Manual operation:
See
8.5.8
Sweep configuration
..................................................................................................88
..................................................................................................88
INITiate<n>:CONTinuous
<State>
This command controls the measurement mode for an individual channel.
Note that in single measurement mode, you can synchronize to the end of the mea-
surement with *OPC, *OPC? or *WAI. In continuous measurement mode, synchroniza-
tion to the end of the measurement is not possible. Thus, it is not recommended that
you use continuous measurement mode in remote control, as results like trace data or
markers are only valid after a single measurement end synchronization.
For details on synchronization see
.
Suffix:
<n>
.
1 | 2
INITiate1 selects single or continuous bargraph measurements.
INITiate2 selects single or continuous scans.
Example:
INIT:CONT OFF
Switches the measurement mode to single measurement.
INIT:CONT ON
Switches the measurement mode to continuous measurement.
Manual operation:
See
See
INITiate<mt>[:IMMediate]
The command initiates a new measurement.