WCDMA TPC Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
821
User Manual 1173.9657.02 ─ 11
Firmware/Software:
V3.2.70
Manual operation:
See
4.5.3.3
Measurement Control Settings
The following commands define measurement control parameters for the TPC mea-
surement.
CONFigure:WCDMa:MEAS<i>:TPC:TOUT
....................................................................... 821
CONFigure:WCDMa:MEAS<i>:TPC:MOEXception
............................................................821
CONFigure:WCDMa:MEAS<i>:TPC:MODE?
.................................................................... 822
CONFigure:WCDMa:MEAS<i>:TPC:MONitor:MLENgth
..................................................... 822
CONFigure:WCDMa:MEAS<i>:TPC:ILPControl:MLENgth?
................................................ 823
CONFigure:WCDMa:MEAS<i>:TPC:ILPControl:AEXecution
...............................................823
CONFigure:WCDMa:MEAS<i>:TPC:ILPControl:TSEF
....................................................... 823
CONFigure:WCDMa:MEAS<i>:TPC:ILPControl:TSGH
.......................................................824
CONFigure:WCDMa:MEAS<i>:TPC:ILPControl:TSSegment
...............................................824
CONFigure:WCDMa:MEAS<i>:TPC:MPEDch:MLENgth
.....................................................824
CONFigure:WCDMa:MEAS<i>:TPC:MPEDch:AEXecution
................................................. 825
CONFigure:WCDMa:MEAS<i>:TPC:ULCM:AEXecution
.....................................................825
CONFigure:WCDMa:MEAS<i>:TPC:ULCM:MLENgth
........................................................ 825
CONFigure:WCDMa:MEAS<i>:TPC:CTFC:MLENgth
.........................................................826
CONFigure:WCDMa:MEAS<i>:TPC:TOUT
<Timeout>
Defines a timeout for the measurement. The timer is started when the measurement is
initiated via a
READ
or
INIT
command. It is not started if the measurement is initiated
manually (ON/OFF key or RESTART/STOP key).
When the measurement has completed the first measurement cycle (first single shot),
the statistical depth is reached and the timer is reset.
If the first measurement cycle has not been completed when the timer expires, the
measurement is stopped. The measurement state changes to
RDY
and the reliability
indicator is set to 1, indicating that a measurement timeout occurred. Still running
READ
,
FETCh
or
CALCulate
commands are completed, returning the available results.
At least for some results there are no values at all or the statistical depth has not been
reached.
A timeout of 0 s corresponds to an infinite measurement timeout.
Parameters:
<Timeout>
Default unit: s
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.1.20
CONFigure:WCDMa:MEAS<i>:TPC:MOEXception
<MeasOnException>
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected.
Command Reference