
WCDMA Multi Evaluation Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
616
User Manual 1173.9657.02 ─ 11
The <carrier> is relevant for the multi carrier configurations. It can be omitted for the
single carrier configuration.
3.5.1.3
FETCh, READ and CALCulate Commands
All commands are used to retrieve measurement results:
●
FETCh...
returns the results of the current measurement cycle (single-shot mea-
surement) after they are valid.
FETCh...
must be used after the measurement
has been started (
INITiate...
, measurement states RUN or RDY).
●
READ...
starts a new single-shot measurement and returns the results.
●
CALCulate...
returns one limit check result per
FETCh
result:
–
OK
: The
FETCh
result is located within the limits or no limit has been defined/
enabled for this result.
–
ULEU
(User limit exceeded upper): An upper limit is violated. The
FETCh
result
is located above the limit.
–
ULEL
(User limit exceeded lower): A lower limit is violated. The
FETCh
result is
located below the limit.
See also: "Retrieving Measurement Results" in the R&S
CMW user manual, chapter
"Remote Control"
3.5.1.4
Current and Statistical Results
The R&S
CMW repeats measurements according to the selected statistic count and
repetition mode. Consecutive measurement values are stored and used to calculate
statistical results, e.g. average, minimum, maximum and standard deviation.
See also: "Statistical Results" in the R&S
CMW user manual, chapter "System Over-
view"
3.5.1.5
Keywords
Selected keywords used in the command description are described in the following.
●
Command usage
If the usage is not explicitly stated, the command allows you to set parameters and
query parameters. Otherwise the command usage is stated as follows:
–
"Setting only": command can only be used to set parameters
–
"Query only": command can only be used to query parameters
–
"Event": command initiates an event
●
Parameter usage
The parameter usage is indicated by the keyword preceding the parameter(s):
–
"Parameters" are sent with a setting or query command and are returned as
the result of a query
–
"Setting parameters" are only sent with a setting command
–
"Query parameters" are only sent with a query command (to refine the query)
–
"Return values" are only returned as the result of a query
Command Reference
深圳德标仪器
135-1095-0799