For a dual carrier scenario with external fading, all parameters described below are
available per carrier.
For general prerequisites, required options and background information
Fig. 2-51: I/Q settings (standard cell fading: external)
................................................................................................ 142
.............................................................................................142
Connector (Out / In)
Selects the output connector. The input connector depends on the output connector
and is displayed for information.
The DIG IQ connectors are located at the rear panel (if an I/Q board is installed).
Remote command:
Sample Rate (Out / In)
The used sample rate is displayed for information. The value is fixed.
Configure the connected instrument accordingly (baseband input settings and digital
I/Q output settings).
Remote command:
Baseband PEP (Out / In)
Indicates the peak envelope power of the baseband signal as dB value relative to full
scale. "Full scale" in this case corresponds to the maximum representable amplitude of
the I/Q samples.
Use the displayed output PEP value to configure the baseband input of the
R&S
AMU200A.