WCDMA Signaling
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
290
User Manual 1173.9657.02 ─ 11
For DAU IMS service: Establishing a voice over IMS call failed.
●
20 (Call Type not Usable)
:
For measurements: The measurement could not be started because the estab-
lished signaling connection had wrong properties.
For DAU IMS service: The voice over IMS settings could not be applied.
●
21 (Call Lost)
:
For measurements: The measurement was interrupted because the signaling con-
nection to the DUT was lost.
For DAU IMS service: The voice over IMS call was lost.
●
23 (Missing Option)
:
The ARB file cannot be played by the GPRF generator due to a missing option.
●
26 (Resource Conflict)
:
The application could not be started or has been stopped due to a conflicting hard-
ware resource or software option that is allocated by another application.
Stop the application that has allocated the conflicting resources and try again.
●
27 (No Sensor Connected)
:
The GPRF External Power Sensor measurement could not be started due to miss-
ing power sensor.
●
30 (File not Found)
:
The specified file could not be found.
●
40 (ARB File CRC Error)
:
The ARB file CRC check failed. The ARB file is corrupt and not reliable.
●
42 (ARB Header Tag Invalid)
:
The ARB file selected in the GPRF generator contains an invalid header tag.
●
43 (ARB Segment Overflow)
:
The number of segments in the multi-segment ARB file is higher than the allowed
maximum.
●
44 (ARB File not Found)
:
The selected ARB file could not be found.
●
45 (ARB Memory Overflow)
:
The ARB file length is greater than the available memory.
●
50 (Startup Error)
:
The Data Application Unit (DAU), a DAU service or a DAU measurement could not
be started. Please execute a DAU selftest.
●
51 (No Reply)
:
The DAU has received no response, for example for a ping request.
●
52 (Connection Error)
:
The DAU could not establish a connection to internal components. Please restart
the instrument.
●
53 (Configuration Error)
:
The current DAU configuration by the user is incomplete or wrong and could not be
applied. Check especially the IP address configuration.
●
54 (Filesystem Error)
:
The hard disk of the DAU is full or corrupt. Please execute a DAU selftest.
●
101 (Firmware Error)
:
Command Reference