ppbRAE 3000 User’s Guide
PID Sensor & Lamp Cleaning/Replacement
The sensor module is made of several components and is attached to
the lamp-housing unit as shown below.
Inlet Probe Assembly
PN 023-3012-000
Sensor Cover
Assembly
Sensor Module Assembly
PN 023-3005-400-FRU
THP (temperature, humidity, pressure) Sensor Module
PN 023-3011-000-FRU
O-ring, 36.5mm x 2.65mm
Stainless Steel Washer
Seal Nut
Lamp 1/2
"
(10.6eV)
PN 050-0000-004
Sensor Detector
PN 023-3010-000
Teflon O-ring
Porous Metal Filter
O-ring, 35mm x 2mm
Filter Fixer
Sensor Components
Note:
The cleaning procedure is not normally needed. Clean the PID
sensor module, the lamp and the lamp housing only when one of the
following has happened:
1.
The reading is inaccurate even after calibration.
2.
The reading is very sensitive to air moisture.
3.
A chemical liquid has been sucked into the unit and damaged
the unit.
Use of the external filter helps to prevent contamination of the sensor.
To access the sensor components and lamp, gently unscrew the lamp-
housing cap, remove the sensor adapter with the gas inlet probe and
the metal filter all together. Then hold the PID sensor and pull it
straight out. A slight, gentle rocking motion helps release the sensor.
93
Содержание ppbRAE 3000
Страница 1: ...ppbRAE 3000 User s Guide Rev C August 2010 P N 059 4021 000 ...
Страница 27: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 25 ...
Страница 45: ...ppbRAE 3000 User s Guide 43 Note Dashed line indicates automatic progression ...
Страница 66: ...ppbRAE 3000 User s Guide 64 Note Dashed line indicates automatic progression ...
Страница 83: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 81 ...
Страница 85: ...ppbRAE 3000 User s Guide 83 Note Dashed line indicates automatic progression ...
Страница 114: ......