ppbRAE 3000 User’s Guide
During duty cycling (PID lamp cleaning), the display shows these
icons in alternation:
If there is a pump failure or obstruction that disrupts the pump, you
will see this icon blinking on and off:
If you see this blinking icon, consult the Troubleshooting section of
this guide.
Calibration Status
The instrument displays this icon if it requires calibration:
Calibration is required (and indicated by this icon) if:
•
The lamp type has been changed (for example, from 10.6 eV
to 9.8 eV).
•
The sensor has been replaced.
•
It has been 30 days or more since the instrument was last
calibrated.
•
If you have changed the calibration gas type without
recalibrating the instrument.