ppbRAE 3000 User’s Guide
Maintenance
The major maintenance items of the instrument are:
•
Battery pack
•
Sensor module
•
PID lamp
•
Sampling pump
•
Inlet connectors and filters
Note: Maintenance should be performed by qualified
personnel only
.
NOTE: The printed circuit board of the instrument is connected
to the battery pack even if the power is turned off
.
Therefore, it is
very important to disconnect the battery pack before servicing or
replacing any components inside the instrument
.
Severe damage
to the printed circuit board or battery may occur if the battery
pack is not disconnected before servicing the unit.
Battery Charging & Replacement
When the display shows a flashing empty battery icon, the battery
requires recharging. It is recommended to recharge the instrument
upon returning from fieldwork. A fully charged battery runs a
instrument for 16 hours continuously. The charging time is less than 8
hours for a fully discharged battery. The battery may be replaced in
the field (in areas known to be non-hazardous), if required.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in area known to be non-hazardous
.
Remove and
replace battery only in areas known to be non-hazardous.
90
Содержание ppbRAE 3000
Страница 1: ...ppbRAE 3000 User s Guide Rev C August 2010 P N 059 4021 000 ...
Страница 27: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 25 ...
Страница 45: ...ppbRAE 3000 User s Guide 43 Note Dashed line indicates automatic progression ...
Страница 66: ...ppbRAE 3000 User s Guide 64 Note Dashed line indicates automatic progression ...
Страница 83: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 81 ...
Страница 85: ...ppbRAE 3000 User s Guide 83 Note Dashed line indicates automatic progression ...
Страница 114: ......