ppbRAE 3000 User’s Guide
Measurement range & resolution
Lamp
Range
Resolution
10.6 eV
1 ppb to 10,000 ppm
1 ppb
9.8 eV
0.01 ppm to 5,000 ppm
10 ppb
11.7 eV
0.01 ppm to 2,000 ppm
10 ppb
Response time (T
90
):
2 seconds
Accuracy
10 to 2000 ppm:
±
3% at calibration point.
(Isobutylene):
PID Detector:
Easy access to lamp and sensor for cleaning
and replacement
Correction Factors:
Over 200 VOC gases built in (based on RAE
Systems Technical Note TN-106)
Calibration:
Two-point field calibration of zero and
standard reference gases
Calibration Reference:
Store up to 8 sets of calibration data, alarm
limits and span values
Inlet Probe:
Flexible 5" tubing
Radio module:
Bluetooth (2.4GHz), RF module (433MHz,
868MHz , 915MHz, or 2.4GHz)
Keypad:
1 operation key and 2 programming keys; 1
flashlight switch
Direct Readout:
Instantaneous, average, STEL, TWA and
peak value, and battery voltage
Intrinsic Safety:
US and Canada: Class I, Division 1, Groups
A, B, C, D
Europe: ATEX (0575 Ex II 2G Ex ia
IIC/IIB T4 Gb)
KEMA 07 ATEX 0127
Complies with EN60079-0:2009,
EN60079-11:2007
12
Содержание ppbRAE 3000
Страница 1: ...ppbRAE 3000 User s Guide Rev C August 2010 P N 059 4021 000 ...
Страница 27: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 25 ...
Страница 45: ...ppbRAE 3000 User s Guide 43 Note Dashed line indicates automatic progression ...
Страница 66: ...ppbRAE 3000 User s Guide 64 Note Dashed line indicates automatic progression ...
Страница 83: ...ppbRAE 3000 User s Guide Note Dashed line indicates automatic progression 81 ...
Страница 85: ...ppbRAE 3000 User s Guide 83 Note Dashed line indicates automatic progression ...
Страница 114: ......