- 91 -
i.e.: the software accepts correctly dosed value within the following thresholds: from
97,500g to 102,500g.
F shortcut keys
As a user you can assign weighing mode functions for quick access. The functions
shall be operated by F1, F2, F3 and F4 buttons.
List of functions to be declared for <DOSING> mode:
NONE
no function assigned to the button
ENTER SAMPLE
function initiates an option of entering the reference
weight along with tolerance (one procedure)
SELECT PRODUCT
function initiates an option of selecting a product from a
product database
LOG IN
function initiates an option of logging in of a selected
user
ENTER TARE
function initiates an option of entering packaging weight
SELECT TARE
function initiates an option of selecting packaging weight,
out of product database
PRINT HEADER
function initiates printout of a designed header
PRINT FOOTER
function initiates printout of a designed footer
VARIABLE 1
function initiates selection and editing of variable 1
VARIABLE 2
function initiates selection and editing of variable 2
Read “F shortcut keys” section of this user manual for procedure on assigning
functions to F buttons.
17.4.1 Entering target weight
Procedure:
•
Press F button to which function <
ENTER SAMPLE
> is assigned, the
editing window is displayed. The target weight value must be given in a
current unit.