- 147 -
Example:
Command:
LDS_1 <CR><LF>
– set last digit option to value ‘always’
Response:
LDS_OK <CR><LF>
– ‘always’ value has been set
Send the latest error code
Format:
ER <CR><LF>
Accessible responses:
ER_A_x <CR><LF>
- command understood, the latest error code given as
a response
ER_I <CR><LF>
- command understood but not accessible at this
moment
x
– error code
Example:
Command:
ER <CR><LF> – give the latest error code
Response:
ER_A_”Err3”
– error code – -Err3- – tarring range exceeded
22.
ERROR MESSAGES
-Err2-
Value beyond zero range
-Err3-
Value beyond tare range
-Err8-
Tarring / Zeroing operation time exceeded
-NULL-
Zero value from converter
-FULL-
Measurement range exceeded
-LH-
Start mass error
Manual number:
IMMU-19-01-07-15-PL
Содержание HRP.R Series
Страница 2: ...2 JULY 2015...
Страница 7: ...7 1 GENERAL INFORMATION Dimensions HRP 16 HRP 32 HRP 62 Dimensions HRP 150...
Страница 8: ...8 1 1 Connectors overview HRP 16 HRP 32 HRP 62 scales HRP 150 scales...
Страница 47: ...47...