- 128 -
20.2
Format of data sent on pressing PRINT button
Printout format
1
2
3
4 - 12
13
14 - 16
17
18
st
ab
ilit
y
m
ar
ker
sp
ace
char
ac
ter
m
ass
sp
ace
uni
t
CR
LF
stability marker - [space] – when stable
- [?] – when unstable
- [^] – if maximum range is exceeded
- [v] – if minimum range is exceeded
character
- [space] – for positive values, or [-] for negative values
mass
- 9 characters – aligned to the right
unit
- 3 characters – aligned to the left.
20.2.1 Format of data sent as response for commands generated from
a computer
After receiving a command, scale responds as follows, first:
XX_A CR LF - command understood and in progress
XX_I CR LF
- command understood but not accessible at this moment
XX_^ CR LF
- command understood but maximum range is exceeded
XX_v CR LF
- command understood but minimum range is exceeded
XX_E CR LF
- an error occurred on command carrying out (time limit exceeded
while waiting for stable measurement result), time limit is scale’s
characteristic parameter
XX
- name of the command
next:
1 – 3
4
5
6
7
8 - 16
17
18 - 20
21
22
com
m
an
d
sp
ace
st
ab
ilit
y
m
ar
ker
sp
ace
char
ac
ter
m
ass
sp
ace
uni
t
CR
LF
command
- 1 ÷ 3 - characters
stability marker - [space] – when stable
- [?] – when unstable
Содержание HRP.R Series
Страница 2: ...2 JULY 2015...
Страница 7: ...7 1 GENERAL INFORMATION Dimensions HRP 16 HRP 32 HRP 62 Dimensions HRP 150...
Страница 8: ...8 1 1 Connectors overview HRP 16 HRP 32 HRP 62 scales HRP 150 scales...
Страница 47: ...47...